1. Must be unidirectional
2. Verified air sweeps over and away from product under dynamic conditions.
3. Velocity profile across entrance plane into work area is required.
4. Acceptable velocity range should be confirmed with airflow smoke pattern test.
5. They are typically 80-100 fpm.
6. Entrance plane is typically 6" below the HEPA filter or diffuser.
Sampling rate of photometer?
Sampling area of probe?
1 CFM
max opening of 1.7 in2 and minimum dimension of 0.5 inches.
Recommended RH and temp?
Use what instruments?
35-60%
Typically 68 or cooler
1% RH
0.1C