Step and shoot (cat) |
Used cables for coupling, 360 rotation clockwise then counter clockwise, table moved per slice |
|
Helical ct |
Continuous acquisition, patient moves steadily |
|
Helical eliminated what |
Cables |
|
What do slip rings allow? |
Helical movement |
|
Anode is made of |
Rhenium, tungsten, molybdenum |
|
Temporal resolution is |
Speed that data can be acquired |
|
Straton improves |
Heat transfer, low cooling rates, gantry rotation |
|
Generator creates.. |
High voltage needed for X-ray production |
|
Generator changes ac (low) to.. |
DC(high)..almost constant potential |
|
Kvp of CT |
80-140 |
|
Ma of CT |
20-500 |
|
High frequency generators are.. |
Small, compact and efficient |
|
Z axis = |
Beam width and slice thickness |
|
Collimators provide constant.. |
Beam thickness(slice thickness) |
|
2 sets of collimators are |
Pre-patient and pre-detector |
|
Prepatient is.. |
Between tube and patient |
|
Decrease arching from.. |
Tungsten deposits |
|
Pre detector is.. |
Below patient, Adove detector |
|
Quicker imaging speed=what to dose |
Decreased |
|
Slip rings use |
A brush like apparatus to provide continuous electrical power and electronica communication from a rotating surface onto a fixed surface |
|
Helical anode is made of |
Graphite alloy |
|
Graphite alloy is.. |
Lighter, high thermal capacity |
|
What is the focal track made of |
Tungsten and rhenium |
|
How long can tubes last for |
150,000-200,000 |
|
High speed rotors are used for.. |
Heat dissipation |
|
Heating unit rating and cooling rate |
8 mil, 1 mhu/minute |
|
Principle cause of CT malfunction |
Tube failure |
|
Straton X-ray tube is newest design which is.. |
Smaller size and anode in oil bath |