- Front
- Back
True or False
If input one is more negative than input 2 the deflection of the waveform will be up |
True
|
|
Impedance is defined as:
|
the resistance against alternating current flow
|
|
A bipolar montage follows which rules?
|
combines serial pairs of electrodes into a chain, in a chain of electrodes input 2 of one channel is the same electrode as input 1 of the next, left side electrodes appear first followed by the right side
|
|
What is the amplitude (uV) of activity that measured 12 mm in height, if a 50 uV calibration signal measured 6 mm?
|
100 uV
|
|
nasopharyngeal electrodes are placed where:
|
through the nasal passage
|
|
If the vertical scale (sensitivity) is 5 uV/mm, then a 60uV will measure:
|
12 mm
(60uV/5uV/mm) |
|
Best montage to measure the amplitude of a sharp wave?
|
A2 ear reference montage
|
|
A faster sweep speed (60mm/sec) allows:
|
easier measurements of faster frequencies, easier ID of 60 hz artifact, better ID of degree of synchrony between 2 hemispheres.
|
|
The use of 60hz notch filter does what:
|
eliminates 60 hz activity only
|
|
The average reference recording is best used for:
|